Publication:
Reducción de riesgos radiológicos potenciales para pacientes sometidos a exámenes de diagnóstico mediante la modificación del espectro de rayos x

dc.contributor.authorDelgado, JSC
dc.contributor.authorPachas, FM
dc.contributor.authorGonzales-Ccoscco, AE
dc.contributor.authorSolís, EC
dc.contributor.authorCastillo, CH
dc.contributor.authorCamargo, GP
dc.contributor.authorAlva-Sánchez, M
dc.date.accessioned2025-02-05T17:29:59Z
dc.date.available2025-02-05T17:29:59Z
dc.date.issued2024
dc.description.abstractThe diagnostic medical practice using X-rays significantly contributes to the collective dose worldwide, where the inherent risk in each examination is proportional to the absorbed dose, which is related to deterministic and stochastic effects of ionizing radiation. Therefore, due to the necessity of optimizing each radiological procedure, this study aimed to reduce the absorbed dose in patients undergoing X-ray examinations by evaluating each parameter that modifies the spectral distribution. The Birch and Marshall method was employed to reconstruct and modify the X-ray spectra based on tube voltage, filtration, tube current, anode angle, and energy pass factor. By modifying these parameters, it was possible to reduce the absorbed dose in the patient’s skin by up to 38%. The proposed methodology is feasible for implementation in clinical centers, given the availability of copper filters incorporated into X-ray equipment. Additionally, an alternative technique involving tantalum filters is presented, achieving a reduction in absorbed dose of up to 57%. Thus, with the developed methodology, it is demonstrated that it is possible to reduce radiation doses by modifying spectral distributions, reproducing medical images with a significant reduction in the absorbed dose on the patient, while ensuring quality and safety in X-ray diagnostic procedures.
dc.formatapplication/pdf
dc.identifier.doihttps: //doi.org/10.15446/mo.n69.115340
dc.identifier.journalMomento
dc.identifier.urihttps://hdl.handle.net/20.500.14703/415
dc.language.isoeng
dc.publisherUniversidad Nacional de Colombia
dc.publisher.countryCO
dc.rightsinfo:eu-repo/semantics/openAccess
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/
dc.subjectabsorbed dose
dc.subjectcopper filters
dc.subjectfiltration
dc.subjectradiographic technique
dc.subjectX-ray spectrum
dc.subject.ocdehttps://purl.org/pe-repo/ocde/ford#3.02.21
dc.titleReducción de riesgos radiológicos potenciales para pacientes sometidos a exámenes de diagnóstico mediante la modificación del espectro de rayos x
dc.typeinfo:eu-repo/semantics/article
dc.type.versioninfo:eu-repo/semantics/publishedVersion
dspace.entity.typePublication

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
115340.pdf
Size:
6.55 MB
Format:
Adobe Portable Document Format